Test Data AC7-01: Difference between revisions
Line 182: | Line 182: | ||
{|align="center" border="1" cellpadding="5" | {|align="center" border="1" cellpadding="5" | ||
|width=600 align=center|<math>{Corr_m=\frac{A_{cm}}{\sum_1^n A_i}}</math>||<math>{(1)}</math> | |width=600 align=center|<math>{Corr_m=\frac{A_{cm}}{\sum_1^n A_i}}</math>||width=50 align=right<math>{(1)}</math> | ||
|- | |- | ||
|<math>{Corr_f=\frac{A_{cf}}{\sum_1^n A_j}}</math>||<math>{(2)}</math> | |<math>{Corr_f=\frac{A_{cf}}{\sum_1^n A_j}}</math>||<math>{(2)}</math> |
Revision as of 17:19, 3 October 2019
Aerosol deposition in the human upper airways
Application Challenge AC7-01 © copyright ERCOFTAC 2019
Test Data
Overview of Tests
Major portions of this section were adopted from Lizal et al. (2015). The positron emission tomography (PET) method provides the best spatial resolution (among radiological methods). In addition to local deposition in the various sections, the deposition hot spots can also be evaluated. However, in comparison to the PET methodology, which is routinely applied to clinical examination, using this method in the in vitro design requires major modifications both in the aerosol preparation and, in particular, in the experiment evaluation approach. The method, based on PET and fulfilling the above mentioned criteria, is presented in the following.
The aerosol exposure procedure
It is a common practice to coat the inner surface of the model, especially when using solid particles, to prevent bouncing of the particles hitting the surface. Since we used liquid di-ethylhexyl sebacate (DEHS) particles, we did not need to coat the inner surface of the model. Another reason for the coating is to prevent surface wetting. In our case the exposure time was short (5 to 15 mins) and only small amounts of DEHS deposited on the walls, therefore the possible flooding of the surface was not an issue. Aerosol particles were generated by a TSI 3475 Condensation Monodisperse Aerosol Generator (CMAG) from TSI, Inc., which works on the controlled heterogeneous condensation principle. Vapours of a suitable material, specifically DEHS, condense by a controlled method on small sodium chloride particles serving as the condensation nuclei. The advantage of DEHS is that it is not hydrophilic and does not evaporate, resulting in a constant size of generated particles. In a standard operating mode, the generator can produce particles with aerodynamic diameters within the 0.1 to 8μm range. The density of DEHS used for the experiments was 0.914 g/cm3 at 25°C. Radioactive aerosol particles were needed for the PET measurement of deposition. Therefore, the solution in the atomizer of the generator had to be tagged by a suitable radioactive substance. Fluorine 18 was the logical Choice of the positron emitter, being easily available at the cooperating PET center and possessing a suitable half-life (109 minutes). A solution of fluorine 18 in the form of fluoride ions was prepared by irradiation of H2180 enriched water on an IBA Cyclone 18/9 cyclotron (irradiation time 25 min, integrated irradiation current 11 μAh) at the UJV Rez's PET Centre in Brno. The irradiated water was transferred by a capillary transport system to a shielded dispensing box, where the fluorine 18 ions were captured on an ion—exchange resin (AG1—X8, BioRad) column and subsequently eluted with 300 ml of 10% sodium Chloride solution, followed by 1Â ml of water for injection. The resulting solution was repeatedly diluted with water for injection until the desired initial radioactivity was achieved. The CMAG was modified for the deposition measurement by using PET so that the atomizer vessel was accommodated in a protective lead container to shield of ionizing radiation. The atomizer was filled with a sodium Chloride solution containing 18F at an initial activity of 2.5 GBq. The concentration of sodium Chloride solution was 20 mg/L. The experimental rig is shown in Figure 5.
Figure 5: A scheme of the experimental setup during the PET measurement of aerosol deposition. |
The generated aerosol was fed through a 85Kr based NEKR,—10 charge equilibrator
(Eckert & Ziegler Cesio) to a PAM aerosol monitor (TSI 3375) for continuous particle
size and concentration measurement. The operation and precision of the aerosol monitor
was validated using Aerodynamic particle sizer (APS) TSI 3321. The validation was
performed prior to the experiment using the identical setup, apart from adding of the
radioactive substance into the atomizer. The size of the particles generated by CMAG was
adjusted according to APS and the size displayed by the aerosol monitor was recorded.
Subsequently, during the experiment, only the radioactive substance was added to the
atomizer and particles of the same size were produced. Usually only a small correction of
saturator flow was needed at the beginning of the experiments. Only one size of particles
was measured during one day; therefore no further adjustments of the generator were
needed. The aerosol monitor served as an on-line indication of the process of aerosol
generation being stable. It was easily accomplished, as the exposure of the model to
the aerosol lasted only 5 to 15 mins. Filters consisting of Millipore AP40 glass fibers
were attached to the output branches of the respiratory tract model. The entire system
(Figure 6)
was enclosed in a plastic bag which was kept in a vacuum to prevent the active
aerosol from leaking into the laboratory. All the 10 terminal branches with flow meters
for flow rate control were combined into one branch with a protective High efficiency
particulate air (HEPA) filter. The vacuum was generated with a Busch R, 5 PA 0008 C
rotary oil vacuum pump. The flow distribution in each section of the model is provided
in Table 3.
Figure 6: A photograph of the physical model prior to the PET measurement. |
The whole exposure of the model to the aerosol was performed in a shielded laboratory
with an underpressure ventilating system, which would prevent the aerosol from escaping
the room in case of a primary safety system failure. The laboratory personnel were not
present in the laboratory during the exposure, with the exception of the regular instrument
supervision. Whenever they had to enter the lab, they wore half mask respirators. The
experiments were performed in a steady—state inhalation mode with the flow rates of 15,
30, and 60 L/min. Liquid monodisperse particles with mass median aerodynamic diameter
of 2.5 and 4.3μm were used. The standard geometric deviation of size was less than 1.24
for all measured regimes. The models were exposed for 10 to 15 minutes depending on
radioactivity decrease by radionuclide decay. The peak activity in the models was 4 to 60
Bq/cm2 (depending on the particle size and concentration, measuring mode, and model
used), as measured with an RP-2000 portable contamination meter (VF Zilina, CZ).
Deposited activity evaluation method
The model was transported to a Siemens Biograph 64 Truepoint PET-CT scanner immediately after the radioactive exposure. The transportation took approximately 3 minutes. The PET-CT scanner acquired firstly CT images, which were promptly followed by PET images. Both the CT and PET images could be attributed unambiguously to the given geometry, owing to the assigned geometrical coordinates. The CT images are important for a precise localization of the edges of the sections, whereas the PET images contain the essential information about aerosol deposition. The CT and PET images were imported into Carimas 2.47 ver. 2.2.44.7002 SW (Turku PET Centre 2012) with the CT images as the main images and the PET images as background images. The software Carimas is a medical image processing tool, which was developed primarily for analysis of PET images at Turku PET Centre in Finland and is available as a freeware for non-commercial use from: http://www.turkupetcentre.fi/carimas/download/. The Carimas was developed in an “interactive data language” (IDL) and runs on IDL Virtual MachineTM, a cross-platform utility for running IDL code. It supports multiple input formats (DICOM, ECAT, Analyse, Interfile, Nifti, Interfile, MicroPET) and general bitmap formats (JPG, TIFF, PNG and BMP). Researchers can perform visualization, segmentation, statistical analysis, or modelling of PET data. It is possible to perform the image fusion, i.e. to coregistrate PET and CT or MR, images. Users can define a volume of interest (VOI) and, by using the static image analysis, they can calculate the mean activity in the VOI (in units of the PET image), standard deviation, minimal and maximal values, the number of voxels analysed, and the volume of the region. The software was tested and compared to commercial tools with an excellent agreement as documented by Nesterov et al. (2012) and Harms et al. (2014). To facilitate the identification of the hot spots in the Carimas software, the BGRY colour system was used to display the PET images and the system range was reduced so that the hot spots were clearly seen in the model sections (Figure 7). The sections of the model were marked as volumes of interest (VOI), and the mean volume activity (in Bq/mL) was evaluated using the Carimas software. Each section was enveloped in an independent VOI. A cylinder or a sphere was selected as the starting shape of the VOI, depending on the shape of the sections. Firstly, the starting shape of the VOI was positioned and shaped as a whole to attain a suitable orientation and an adequate size. In the next step, each VOI was shaped in the vertex mode for a local mesh adjustment, so that the VOI surrounded the section with an overreach, while preventing an overlap of different VOIs. If the latter was not avoided, the observed activity would be attributed to both VOIs, resulting in biased results. The degree of mesh density can be modified by reducing the number of nodes or by enlarging or shrinking the mesh.
Figure 7: CT and PET image presentation in the Carimas software. |
With regard to the PET precision, a VOI should optimally overreach the section by
roughly 5mm on all sides to cover all the radiation emitted from the section. However, due
to the complex lung geometry, it was impossible to form all the VOIs with such a large
overreach, and so the uncounted radioactivity was accounted for in a correction VOI. Two
additional VOIs had to be created to determine the magnitude of the correction: a VOI for
the top part of the box, accommodating the model with the hoses, and a correction VOI
for the bottom part of the box, accommodating the filters downstream of the terminal
branches. They encompass the large top space with the model and the bottom space
between the two horizontal partitions respectively in Figure 6.
Two separate corrections
were needed for the model and filters, because the difference in volume radioactivity is
within few orders of magnitude and therefore it is necessary to assign the uncounted
radioactivity to its real source. It is essential to preserve the true ratio of the radiation
deposited in model sections and on filters. The correction factor for the sections was
calculated as the ratio of the total activity obtained from the VOI encompassing the
whole model to the activity obtained by summing up the section VOIs. Similarly, the
correction factor for the filters was calculated as the ratio of the total activity of the large
VOI encompassing all the filters, to the sum of the activities of the filter VOIs. The
corrections for the model sections
() and for the filters () were calculated as
follows:
width=50 align=right | |
where and are the activities measured in the model
sections and on the filters respectively, and and
are the activities measured in the correction VOI for the
model and filters respectively. We assume that the distribution of uncounted radioactivity
is proportional to the measured activity in the sections of the model. The higher the
activity measured in the filter, the higher the activity spread around the section.
Contributed by: *** — ***
© copyright ERCOFTAC 2019